1. Introduction
2. Detector fabrication
Fig. 1. (a) Optical image of a 4 × 4 TES μ-calorimeter array (b) Design layout of a single TES pixel (c) The cutaway view of a single TES pixel. |
Table 1. Geometrical parameters of the μ-calorimeter pixel. |
| TES detector | Absorber | SiNx membrane | Supporting pillars | |
|---|---|---|---|---|
| Dimensions | 150 μm × 150 μm | 190 μm × 190 μm | 210 μm × 210 μm | 5 μm × 5 μm (small pillar) 20 μm × 20 μm (center pillar) |
| Thickness | 425 nm (Mo45/Au30/Au350) | 2.5 μm | 500 nm | 1.8 μm |
3. Electrothermal parameters of the μ-calorimeter
3.1. Transition temperature (TC)
Fig. 2. (a) R-T curve of the μ-calorimeter under a small excitation current of 10μA, TC(onset) is approximately 91.6mK. (b) I-V curves of the μ-calorimeter at various bath temperature. |
3.2. I-V characteristics
3.3. Thermal conductance (G) and heat capacity (C)
Fig. 3. (a) Joule Power of the Mo/Au/Au μ-calorimeter with R0 = 0.8 Rn versus bath temperature. (b) The signal pulse of the μ-calorimeter measured at a bath temperature close to the superconducting transition temperature of the detector. |
Table 2. Summary of the electro-thermal parameters. |
| Tc(onset) = 91.6mK | Rn = 17.5mΩ | C = 0.737pJ/K |
|---|---|---|
| n = 3.05 | T0.8 = 90.3mK | G(T0.8) = 319pW/K |
| Parameters | Mathematical Expression | Value |
| I0 | N/A | 32.9 μA |
| R0 | N/A | 6.26 mΩ |
| $T_0$ | N/A | 86.5 mK |
| $G(T_0)$ | $\begin{array}{l} \mathrm{G}\left(\mathrm{T}_{0}\right)=\frac{\mathrm{dP}_{\text {both }}}{\mathrm{dT}}=\mathrm{nKT} \mathrm{T}^{\mathrm{n}-1}= \\ \mathrm{G}\left(\mathrm{T}_{0.8}\right) *\left(\frac{\mathrm{T}}{\mathrm{T}_{0.8}}\right)^{\mathrm{n}-1} \end{array}$ | 292 pW/K |
| $α_I$ | $\alpha_{I}=\left.\frac{T_{0}}{R_{0}} \frac{\partial R}{\partial T}\right|_{I_{0}}$ | 47.2 |
| $β_I$ | $\beta_{I}=\left.\frac{I_{0}}{R_{0}} \frac{\partial R}{\partial I}\right|_{T_{0}}$ | 1.15 |
| τ | $\tau=\frac{\mathrm{C}}{\mathrm{G}}$ | 2.31 ms |
| $L_I$ | $\mathscr{L}_{I}=\frac{\alpha_{I} P_{0}}{G\left(T_{0}\right) T_{0}}$ | 11.62 |
| $τ_I$ | $\tau_{I}=\frac{\tau}{1-\mathscr{L}_{I}}$ | −217 μs |
| $τ_+$ | $\tau_{+}=\tau_{e l} \approx \frac{L}{R_{0}\left(1+\beta_{I}\right)}$ | 4.59 μs |
| $τ_-$ | $\tau_{-}=\tau_{e f f} \approx \tau \frac{1+\beta_{I}}{1+\beta_{I}+\mathscr{L}_{I}}$ | 361 μs |
3.4. Temperature sensitivity (α) and current sensitivity (β)
Fig. 4. (a) Temperature sensitivity, (b) current sensitivity of the Mo/Au/Au μ-calorimeter measured at various working points. |
3.5. Summary of electrothermal parameters
4. Noise analysis and estimation of the energy resolution
Fig. 5. (a) Current noise spectrum of the μ-calorimeter. (b) the amplitude of power to current responsivity versus frequency. |
5. μ-calorimeter response to X-ray photons
Fig. 6. Schematic view of the X-ray response measurement setup. |
Fig. 7. (a) A typical current pulse of the TES μ-calorimeter. The red line was the double exponential fitting of the raw pulse. The inset is the calibration curve of TES for to translate the pulse amplitude into photon energy. (b)the measured energy spectrum of Mn Kα by Mo/Au/Au TES. The red solid line is used to fit the data with Eq. (5.3). The blue dash line represents the natural line shape of Mn Kα lines. The inset gives the energy resolution measured at six different working points. |

