摘要:
目的:评估迭代去金属伪影(iterative metal artifact reduction,iMAR)技术在去除腹部金属植入物伪影对CT图像质量改善中的应用价值。方法:回顾性收集32例有腹部金属植入物的受检者的CT图像,分别采用iMAR和滤波反向投影(filtered back projection,FBP)2种算法进行重建,由2名放射科医师对图像进行主观质量评分,比较2种算法获得图像间的质量评分差异。测量2种算法处理后金属植入物附近伪影最严重层面的噪声值,计算图像信噪比(signal-to-noise ratio,SNR)、对比度噪声比(contrast-to-noise ratio,CNR)和伪影指数(artifact index,AI),并比较2种算法间以上指标差异及测量间的一致性。结果:iMAR算法重建图像的主观质量评分明显高于FBP算法,差异有统计学意义(S=232.500,P<0.050)。iMAR重建图像的SNR及CNR明显高于FBP重建图像,差异有统计学意义(SNR,1.890±1.864比0.720±1.384, P=0.002; CNR,35.872±13.154比15.245±12.362, P<0.050);iMAR重建图像的AI明显低于FBP重建图像,差异有统计学意义(32.690±15.694比152.594±117.203, P<0.050)。结论:iMAR技术能够明显减少腹部植入物金属伪影对图像的影响,可改善图像质量,具有良好的临床应用价值。
中图分类号:
周慧赟, 吕晓宇, 李纪强, 董海鹏, 张伟华. iMAR在腹部金属植入物伪影去除中的应用价值[J]. 诊断学理论与实践, 2022, 21(03): 380-384.
ZHOU Huiyun, LÜ Xiaoyu, LI Jiqiang, DONG Haipeng, ZHANG Weihua. Application value of iMAR algorithm in reducing metal artifacts from abdominal implants[J]. Journal of Diagnostics Concepts & Practice, 2022, 21(03): 380-384.
[1] |
Morsbach F, Bickelhaupt S, Wanner GA, et al. Reduction of metal artifacts from hip prostheses on CT images of the pelvis: value of iterative reconstructions[J]. Radiology, 2013, 268(1):237-244.
doi: 10.1148/radiol.13122089 pmid: 23513244 |
[2] |
Boas FE, Fleischmann D. Evaluation of two iterative techniques for reducing metal artifacts in computed tomography[J]. Radiology, 2011, 259(3):894-902.
doi: 10.1148/radiol.11101782 URL |
[3] |
Lee MJ, Kim S, Lee SA, et al. Overcoming artifacts from metallic orthopedic implants at high-field-strength MR imaging and multi-detector CT[J]. Radiographics, 2007, 27(3):791-803.
doi: 10.1148/rg.273065087 URL |
[4] | 李杰, 孙兴文, 袁慧书. 去金属伪影技术的发展应用[J]. 中华放射学杂志, 2021, 55(9):889-894. |
Li J, Sun XW, Yuan HS. The development and application of metal artifact reduction techniques[J]. Chin J Radiol, 2021, 55(9):889-894. | |
[5] |
Pua R, Wi S, Park M, et al. An image-based reduction of metal artifacts in computed tomography[J]. J Comput Assist Tomogr, 2016, 40(1):131-141.
doi: 10.1097/RCT.0000000000000316 URL |
[6] |
Axente M, Paidi A, Von Eyben R, et al. Clinical evaluation of the iterative metal artifact reduction algorithm for CT simulation in radiotherapy[J]. Med Phys, 2015, 42(3):1170-1183.
doi: 10.1118/1.4906245 URL |
[7] |
Hu Y, Pan S, Zhao X, et al. Value and clinical application of orthopedic metal artifact reduction algorithm in CT scans after orthopedic metal implantation[J]. Korean J Radiol, 2017, 18(3):526-535.
doi: 10.3348/kjr.2017.18.3.526 URL |
[8] |
Wellenberg RH, Boomsma MF, van Osch JA, et al. Computed tomography imaging of a hip prosthesis using ite-rative model-based reconstruction and orthopaedic metal artefact reduction: a quantitative analysis[J]. J Comput Assist Tomogr, 2016, 40(6):971-978.
pmid: 27331924 |
[9] |
Laurent G, Villani N, Hossu G, et al. Full model-based iterative reconstruction (MBIR) in abdominal CT increa-ses objective image quality, but decreases subjective acceptance[J]. Eur Radiol, 2019, 29(8):4016-4025.
doi: 10.1007/s00330-018-5988-8 URL |
[10] |
Ishikawa T, Suzuki S, Harashima S, et al. Metal artifacts reduction in computed tomography: a phantom study to compare the effectiveness of metal artifact reduction algorithm, model-based iterative reconstruction, and virtual monochromatic imaging[J]. Medicine (Baltimore), 2020, 99(50):e23692.
doi: 10.1097/MD.0000000000023692 URL |
[11] | 刘颖, 张磊, 谢丽响, 等. O-MAR技术去金属植入物伪影的临床价值[J]. 中国医学计算机成像杂志, 2018, 24(1):92-95. |
Liu Y, Zhang L, Xie LX, et al. Application of O-MAR technique in reducing the metallic implants artifacts[J]. Chin Comp Med Imaging, 2018, 24(1):92-95. | |
[12] | 郁斌, 吕富荣, 张黎, 等. 胸部CT扫描中迭代去金属伪影算法减少金属伪影的有效性[J]. 中国医学影像技术, 2017, 33(4):590-593. |
Yu B, Lv FR, Zhang L, et al. Effectiveness of iterative metal artifact reduction for reduction of metal artifact in chest CT scanning[J]. Chin J Med Imaging Technol, 2017, 33(4):590-593. | |
[13] |
Wang F, Xue H, Yang X, et al. Reduction of metal artifacts from alloy hip prostheses in computer tomography[J]. J Comput Assist Tomogr, 2014, 38(6):828-833.
doi: 10.1097/RCT.0000000000000125 URL |
[14] | 李传东, 陈晓亮, 赵其超, 等. 基于SEMAR技术在CT检查中去除金属假体伪影临床应用价值的研究[J]. 临床放射学杂志, 2017, 36(7):1038-1041. |
Li CD, Chen XL, Zhao QC, et al. The clinical significance of the metal arthroplasty artifact reduction based on SEMAR technology[J]. J Clin Radiol, 2017, 36(7):1038-1041. | |
[15] | 姚开情, 吕发金, 吕亮, 等. IMAR在去除髋关节植入物金属伪影中的临床价值[J]. 重庆医学, 2020, 49(12):1943-1945. |
Yao KQ, Lv FJ, Lv L, et al. Clinical value of iterative metal artifact reduction in reducing metal artifacts of hip implantation material[J]. Chongqing Med, 2020, 49(12):1943-1945. | |
[16] | Wayer DR, Kim NY, Otto BJ, et al. Unintended consequences: review of new artifacts introduced by iterative reconstruction CT metal artifact reduction in spine ima-ging[J]. AJNR, 2019, 40(11):1973-1975. |
[17] |
Kotsenas AL, Michalak GJ, DeLone DR, et al. CT metal artifact reduction in the Spine: Can an iterative reconstruction technique improve visualization?[J]. AJNR, 2015, 36(11):2184-2190.
doi: 10.3174/ajnr.A4416 URL |
[18] |
Wellenberg RHH, van Osch JAC, Boelhouwers HJ, et al. CT radiation dose reduction in patients with total hip arthroplasties using model-based iterative reconstruction and orthopaedic metal artefact reduction[J]. Skeletal Radiol, 2019, 48(11):1775-1785.
doi: 10.1007/s00256-019-03206-z pmid: 31016340 |
[19] |
Hakim A, Pastore-Wapp M, Vulcu S, et al. Efficiency of iterative metal artifact reduction algorithm (iMAR) applied to brain volume perfusion CT in the follow-up of patients after coiling or clipping of ruptured brain aneurysms[J]. Sci Rep, 2019, 9(1):19423.
doi: 10.1038/s41598-019-55792-6 URL |
[20] |
Shinohara Y, Ohmura T, Sasaki F, et al. Appropriate iMAR presets for metal artifact reduction from surgical clips and titanium burr hole covers on postoperative non-contrast brain CT[J]. Eur J Radiol, 2021, 141:109811.
doi: 10.1016/j.ejrad.2021.109811 URL |
No related articles found! |
阅读次数 | ||||||
全文 |
|
|||||
摘要 |
|
|||||