Performance verification of a full automatic enzyme immunoassay analyzer for the detecting of HAV-IgM antibody

WANG Fenglei1, WU Xiaoyan1(), WANG Zhicheng2   

  1. 1. Department of Laboratory, Qingpu Hospital of Traditional Chinese Medicine, Shanghai 201799, China
    2. Department of Clinical Laboratory, Huashan Hospital Affiliated to Fudan University, Shanghai 200040, China
  • Received:2019-11-20 Online:2020-02-25 Published:2020-02-25
  • Contact: WU Xiaoyan E-mail:qher@163.com

Abstract:

Objective: To verify and evaluate the performance of a full automatic enzyme immunoassay analyzer (Tecan) for the detecting of hepatitis A virus(HAV)-IgM. Methods: According to the EP5-A2 document of Clinical and Laboratory Standards Institute(CLSI)and related documents of China National Accreditation Service for Conformity Assessment-CL39(CNAL-CL39) in combination with the condition of our laboratory, the precision, detection limit, accuracy and cut off value of the full automatic enzyme immunoassay analyzer Tecan for the detecting of HAV-IgM were eva-luated.Results: The intra-and inter-batch coefficient of variation (CV) of Tecan for detecting HAV-IgM were 7.77% and 11.17%, respectively, and the coincidence rates of both positive and negative results were 100%. The detection limit was 1.0 NCU/mL, which met the requirement of laboratory and was close to the statement of manufacturer (2.0 NCU/mL). In the validation experiment for cut-off value, the light absorption value ($\bar{x}$+3SD)=0.032 was less than the value 0.105 provided by the kit. Conclusions: The precision, accuracy, detection limit and cutoff value of the automatic enzyme immunoassay analyzer Tecan for detecting HAV-IgM meet the requirements of manufacturer's standards and clinical needs.

Key words: Hepatitis A virus (HAV)-IgM, Full automatic enzyme immunoassay analyzer, Performance verification