The impact of different diluting media on determination of serum estradiol concentration by using chemiluminescent immunoassay
Received date: 2018-12-20
Online published: 2019-08-25
Objective: To evaluate the impact of different diluting media on determination of serum estradiol (E2) concentration by using chemiluminescent immunoassay. Methods: Fifty serum samples were collected for E2 detection using chemiluminescent immunoassay by an automatic biochemical analyzer (UniCel DXI 800, Beckman Coulter). Four diluting media were used, including standard diluent S0 (Beckman Coulter), pooled low-level E2 serum (PS), E2/progesterone standard diluent (Roche), and calf serum (CS). Samples were manually diluted into 1∶1 or 1∶4 using the above 4 media with two parallel tubes and detected by UniCel DXI 800. Multiple statistical methods were applied to analyze the data. Results: Students't test revealed that the E2 concentration measured by diluting media CS, Roche and PS compared with S0 showed significant differences statistically (P=5.03×10-11, 1.44×10-10 and 2.79×10-13, respectively). Pearson correlation analysis showed that the correlation coefficient r between diluting media S0 and CS, Roche, PS were 0.992 3, 0.971 3, 0.968 0, respectively. After correction with linear regression model, the concentration measured by diluting media CS, Roche and PS compared with S0 showed no significant difference statistically (P=0.998 3, 0.997 9, 0.987 7, respectively). Consistence of E2 concentration measured by diluting media CS, Roche and PS compared with S0 was evaluated by Bland-Altman method, the results showed that 2%, 2% and 8% of the samples exceeded the consistence limit, respectively. The maximum bias of three assessed diluting media were 16.8%, 44.8% and 45.2%, respectively. CS shows the best consistence. Conclusions: Among the three assessed diluting media, CS shows the best performance and could serve as a replacement of manufacturer's standard diluent with high stability and lower cost.
Key words: Estradiol; Diluent; Chemiluminescent immunoassay; Bland-Altman method
SI Chenchen, XU Bufang, FAN Weimin, ZHANG Aijun . The impact of different diluting media on determination of serum estradiol concentration by using chemiluminescent immunoassay[J]. Journal of Diagnostics Concepts & Practice, 2019 , 18(04) : 448 -453 . DOI: 10.16150/j.1671-2870.2019.04.013
[1] | 李茹, 金彩卿. 血清雌二醇与孕酮检测在诊断先兆流产及预测先兆流产孕妇妊娠结局方面的应用价值[J]. 当代医药论丛, 2019, 17(2):92-94. |
[2] | 冼碧茹, 李玉香. 血清雌二醇和孕酮水平预测体外受精行长效激动剂方案妊娠结局的研究[J]. 实用妇产科杂志, 2018, 34(5): p.366-370. |
[3] | 张春燕, 潘柏申. 激素检测在女性不孕症诊治中的临床应用[J]. 检验医学, 2015, 30(7):761-765. |
[4] | 普天磊, 薛嫚, 沈慈丹, 等. 血清雌二醇水平分析方法研究进展[J]. 生殖与避孕, 2016, 36(12):1017-1024. |
[5] | 包安裕, 李艳, 苏汉文. 稀释介质与稀释方式对化学发光免疫法测定血清雌二醇浓度的影响[J]. 标记免疫分析与临床, 2011, 18(4):271-273. |
[6] | 李洋, 何瑞冰, 曹振毅, 等. 不同稀释介质及方式对化学发光免疫法测定血清雌二醇浓度的影响[J]. 临床输血与检验, 2015, 17(4):327-330. |
[7] | 王洪珍, 陈小芳. 不同稀释液对化学发光免疫法测定血清雌二醇浓度的影响[J]. 国际检验医学杂志, 2016, 37(13):1827-1828. |
[8] | Tholen DW. Evaluation of linearity in the clinical laboratory[J]. Arch Pathol Lab Med, 2004, 128(10):1078. |
[9] | Tholen DW, Musiol-Kroll EM, Astles JR, et al. Evaluation of the linearity of quantitative measurement procedures: a statistical approach; approved guideline[R/OL]. 2003[2018-12-20]. https://www.scienceopen.com/document?vid=13d8c337-d7d6-41ca-9762-ac422c22bc00. |
[10] | 牛志宏, 冯云. 血清性激素监测在促排卵周期中的作用[J]. 生殖医学杂志, 2013, 22(10):774-778. |
[11] | 牛志宏, 冯云, 黄晓燕, 等. 体外受精-胚胎移植周期雌二醇下降对助孕的影响[J]. 上海交通大学学报(医学版), 2006, 26(12):1399-1401. |
[12] | 王佳, 蔡新. UniCel DXI800全自动微粒子化学发光免疫分析系统性能评价[J]. 临床检验杂志, 2008, 26(2):157-158. |
[13] | 肖翔, 张伟龙, 潘理想, 等. UniCel DXI800全自动磁性微粒子化学发光免疫分析系统性能评价[J]. 检验医学与临床, 2013, 10(2):140-141. |
[14] | 高云朝, 杨亚东, 陆汉魁. 稀释介质对化学发光免疫分析检测血清HCG的影响[J]. 标记免疫分析与临床, 2007, 14(2):121,72. |
[15] | 卢德权. 不同稀释液对β-HCG稀释检测的影响[J]. 泰山医学院学报, 2014, 35(8):744-746. |
[16] | 陈卉. Bland-Altman分析在临床测量方法一致性评价中的应用[J]. 中国卫生统计, 2007, 24(3):308-309,315. |
[17] | 李镒冲, 李晓松. 两种测量方法定量测量结果的一致性评价[J]. 现代预防医学, 2007, 34(17):3263-3266,3269. |
[18] | 吕世静. 临床免疫学检验[M]. 中国医药科技出版社, 2015. |
[19] | Cole LA, Rinne KM, Shahabi S, et al. False-positive hCG assay results leading to unnecessary surgery and chemotherapy and needless occurrences of diabetes and coma[J]. Clin Chem, 1999, 45(2):313-314. |
[20] | Kricka LJ. Interferences in immunoassay--still a threat[J]. Clin Chem, 2000, 46(8 Pt 1):1037-1038. |
[21] | 韦维, 戴盛明. 异嗜性抗体在免疫测定中干扰的研究进展[J]. 国际检验医学杂志, 2010, 31(10):1123-1126. |
[22] | 蒋利君, 黎宇, 戴盛明. 异嗜性抗体对免疫测定干扰的研究进展[J]. 分子诊断与治疗杂志, 2010, 2(1):68-72. |
[23] | 秦川. 实验动物学[M]. 人民卫生出版社: 2010. |
[24] | 宋立荣, 李思经. 谈新生牛血清的质量控制[J]. 黑龙江畜牧兽医, 2008, 51(10):92-94. |
/
〈 |
|
〉 |